Найти
Телефон
Меню

Официальный представитель завода АО «Таурас-Феникс»

в Уфе, в Казани, в Оренбурге, в Перми, в Ижевске и в Казахстане

Taures-fenix
+7 347 201 77 10 Бесплатно по России

PI2209 – фото
Фото
Видео

На PI2209

Для заявки пишите в МАХ или Telegram

# #

Характеристики

Output signal switching signal; analog signal; IO-Link; (configurable)
Measuring range -100...100
Medium temperature -20...200; (-20...90 < 0 bar)
Electrical design PNP/NPN
Process connection Clamp DN40 (1,5")

Все характеристики

Number of inputs and outputs Number of digital outputs: 2; Number of analog outputs: 1
Measuring range
-1000...1000 mbar-14.5...14.5 psi-100...100 kPa
Process connection Clamp DN40 (1,5") DIN 32676 (ISO 2852)
System gold-plated contacts
Application flush mountable for the food and beverage industry
Media viscous media and liquids with suspended particles; liquids and gases
Medium temperature -20...200; (-20...90 < 0 bar)
Min. bursting pressure
30000 mbar435 psi3000 kPa
Pressure rating
10000 mbar145 psi1000 kPa
Type of pressure relative pressure; vacuum
Min. insulation resistance 100; (500 V DC)
Protection class III
Reverse polarity protection yes
Integrated watchdog yes
Step response time analog output 45
Min. response time of switching output (dAP) 3
Step response time analog output 7
Number of inputs and outputs Number of digital outputs: 2; Number of analog outputs: 1
Total number of outputs 2
Output signal switching signal; analog signal; IO-Link; (configurable)
Electrical design PNP/NPN
Number of digital outputs 2
Output function normally open / closed; (configurable)
Number of analog outputs 1
Analog current output 4...20, invertible; (scalable)
Short-circuit protection yes
Type of short-circuit protection yes (non-latching)
Overload protection yes
Measuring range
-1000...1000 mbar-14.5...14.5 psi-100...100 kPa
Set point SP
-998...1000 mbar-14.45...14.5 psi-99.8...100 kPa
Reset point rP
-1000...998 mbar-14.5...14.45 psi-100...99.8 kPa
Analog start point
-1000...500 mbar-14.5...7.25 psi-100...50 kPa
Analog end point
-500...1000 mbar-7.25...14.5 psi-50...100 kPa
In steps of
1 mbar0.05 psi0.1 kPa
Factory setting
SP1 = -500 mbarrP1 = -540 mbar
SP2 = 500 mbarrP2 = 460 mbar
ASP = -1000 mbarAEP = 1000 mbar
Switch point accuracy < ± 0,2; (Turn down 1:1)
Repeatability < ± 0,1; (with temperature fluctuations < 10 K; Turn down 1:1)
Characteristics deviation < ± 0,2; (Turn down 1:1 , linearity, incl. hysteresis and repeatability , limit value setting to DIN EN IEC 62828-1)
Linearity deviation < ± 0,15; (Turn down 1:1)
Hysteresis deviation < ± 0,15; (Turn down 1:1)
Long-term stability < ± 0,1; (Turn down 1:1; per year)
Temperature coefficient zero point < ± 0,3; (0...200 °C)
Temperature coefficient span < ± 0,15; (0...200 °C)
Damping process value dAP 0...30
Damping for the analog output dAA 0.01...99.99
Communication interface IO-Link
Transmission type COM2 (38,4 kBaud)
IO-Link revision 1.0
SDCI standard IO-Link Communication Specification, Version 1.0, January 2009, Ordner No: 10.002
SIO mode yes
Required master port class A
Process data analog 1
Process data binary 2
Min. process cycle time 2.3
Supported DeviceIDs
Type of operationDeviceID
default258
Ambient temperature -25...65
Note on ambient temperature
Medium temperature: ≤ 200 °C
Storage temperature -25...100
Protection IP 67; IP 68; IP 69K
EMC
DIN EN 61000-6-2
DIN EN 61000-6-3
Shock resistance
DIN EN 60068-2-2750 g (11 ms)
Vibration resistance
DIN EN 60068-2-620 g (10...2000 Hz)
MTTF 141
Note on approval
Including free 6-point calibration certificate.
Weight 1074.4
Housing tubular
Dimensions Ø 50 / L = 163
Material stainless steel (1.4404 / 316L); stainless steel (1.4435 / 316L); PBT; PEI; PFA
Materials (wetted parts) stainless steel (1.4435 / 316L)
Min. pressure cycles 10 million
Process connection Clamp DN40 (1,5") DIN 32676 (ISO 2852)
Surface characteristics Ra/Rz of the wetted parts Ra < 0.38 µm
Fill fluid for pressure transfer NEOBEE (food grade, FDA compliant)
Diaphragm diameter 34
Display
Display unitLED, green
Switching statusLED, yellow
Function displayalphanumeric display, 4-digit
Measured valuesalphanumeric display, 4-digit
Display unit mbar; kPa; psi; inH2O; mWS; % of the span
Pack quantity 1 pcs.
Connector: 1 x M12; coding: A; Contacts: gold-plated

Приемущества

For use in high temperatures in the food and beverage industry
Robust stainless steel housing with hygienic front-flush design
With two-wire and three-wire connection technology
Reduced wiring complexity when operated as two-wire current loop
Integrated clamp diaphragm seal with stainless steel membrane
Convenient operation via buttons or parameter setting via IO-Link