Характеристики

Output signal switching signal; analog signal; IO-Link
Medium temperature -25...150
Measuring range -50...1004
Electrical design PNP/NPN
Process connection G 1 external thread sealing cone

Все характеристики

Number of inputs and outputs Number of digital outputs: 2; Number of analog outputs: 1
Measuring range
-0.124...2.5 bar-1.8...36.25 psi-50...1004 inH2O-12.4...250 kPa
Process connection threaded connection G 1 external thread sealing cone Attention: The unit must only be installed in a process connection for G1 sealing cone.; The G1A sealing cone of the unit is only suited for adapters with metal end stop.
System gold-plated contacts
Application flush mountable for the food and beverage industry
Media viscous media and liquids with suspended particles; liquids and gases
Medium temperature -25...150
Min. bursting pressure
50 bar725 psi5000 MPa
Pressure rating
20 bar290 psi2000 kPa
Vacuum resistance
-1000 mbar-0.1 MPa
Type of pressure relative pressure; vacuum
No dead space yes
MAWP (for applications according to CRN) 20
Min. insulation resistance 100; (500 V DC)
Protection class III
Reverse polarity protection yes
Integrated watchdog yes
Step response time analog output 30
Min. response time of switching output (dAP) 3
Step response time analog output 7
Number of inputs and outputs Number of digital outputs: 2; Number of analog outputs: 1
Total number of outputs 2
Output signal switching signal; analog signal; IO-Link
Electrical design PNP/NPN
Number of digital outputs 2
Output function normally open / closed; (configurable)
Number of analog outputs 1
Analog current output 4...20, invertible; (scalable)
Short-circuit protection yes
Type of short-circuit protection yes (non-latching)
Overload protection yes
Measuring range
-0.124...2.5 bar-1.8...36.25 psi-50...1004 inH2O-12.4...250 kPa
Set point SP
-0.12...2.5 bar-1.74...36.26 psi-48...1004 inH2O-12...250 kPa
Reset point rP
-0.124...2.496 bar-1.8...36.2 psi-50...1002 inH2O-12.4...249.6 kPa
Analog start point
-0.124...1.994 bar-1.8...28.92 psi-50...801 inH2O-12.4...199.4 kPa
Analog end point
0.382...2.5 bar5.54...36.26 psi153...1004 inH2O38.2...250 kPa
Min. difference between SP and rP
0.004 bar0.06 psi2 inH2O0.4 kPa
In steps of
0.001 bar0.01 psi1 inH2O0.1 kPa
Factory setting
SP1 = 0.625 barrP1 = 0.575 bar
SP2 = 1.875 barrP2 = 1.825 bar
ASP = 0.00 barAEP = 2.50 bar
dAP = 2.00 sdAA =2.00 s
Accuracy ± 2,5+ (0,08 x ( Umgebungstemperatur - Mediumtemperatur ))
Repeatability ± 0,2
Resolution 0.2
Switch point accuracy < ± 0,2; (DIN EN IEC 62828-1; Turn down 1:1)
Repeatability < ± 0,1; (with temperature fluctuations < 10 K; Turn down 1:1)
Characteristics deviation < ± 0,2; (DIN IEC EN 62828-1 incl. zero point and span error, non-linearity, hysteresis; Turn down 1:1)
Linearity deviation < ± 0,15; (Turn down 1:1)
Hysteresis deviation < ± 0,15; (Turn down 1:1)
Long-term stability < ± 0,1; (Turn down 1:1; per year)
Total deviation over temperature range
Temperature rangetotal deviation
-25...15 °CCharacteristics deviation ± 0,05 % of the span / 10 K
15...80 °CCharacteristics deviation
80...150 °CCharacteristics deviation ± 0,1 % of the span / 10 K
Notes on the accuracy / deviation for further details see section Diagrams and graphs
Damping process value dAP 0...99.99
Damping for the analog output dAA 0...99.99
Communication interface IO-Link
Transmission type COM2 (38,4 kBaud)
IO-Link revision 1.1
SDCI standard IEC 61131-9
Profiles
Common - I&DIdentification and Diagnosis
FunctionMeasurement data, high resolution
SIO mode yes
Required master port class A
Min. process cycle time 5.6
IO-Link resolution pressure 0.0001
IO-Link resolution temperature 0.2
IO-Link process data (cyclical)
Functionbit length
pressure32
temperature32
device status4
binary switching information2
IO-Link functions (acyclical) application specific tag; internal temperature; operating hours counter; switching cycles counter; Pressure peak counter
Supported DeviceIDs
Type of operationDeviceID
default1154
Ambient temperature -25...80
Storage temperature -40...100
Protection IP 67; IP 68; IP 69K
EMC
DIN EN 61326-1
Shock resistance
DIN EN 60068-2-2750 g (11 ms)
Vibration resistance
DIN EN 60068-2-620 g (10...2000 Hz)
MTTF 214
Note on approval
Factory certificate available as download at www.factory-certificate.ifm
UL approval
UL approval numberJ049
File number ULE174189
Weight 384.8
Housing tubular
Dimensions Ø 50 / L = 106
Material stainless steel (1.4404 / 316L); FKM; PTFE; PBT; PEI; PFA
Materials (wetted parts) ceramics (99.9 % Al2O3); stainless steel (1.4435 / 316L) surface characteristics: Ra < 0,4 µm / Rz = 4 µm; PTFE
Min. pressure cycles 100 million
Tightening torque 20
Process connection threaded connection G 1 external thread sealing cone Attention: The unit must only be installed in a process connection for G1 sealing cone.; The G1A sealing cone of the unit is only suited for adapters with metal end stop.
Display
Display unitLED, green
Switching statusLED, yellow
Function displayalphanumeric display, 4-digit
Measured valuesalphanumeric display, 4-digit
Display unit bar; psi; kPa; inH2O
Pack quantity 1 pcs.
Connection Connector: 1 x M12; coding: A; Contacts: gold-plated

Приемущества

Robust stainless steel design for the requirements of the food and beverage industry
Hygienic, flush design
Resistant to high-pressure cleaning with aggressive cleaning agents
High temperature resistance and high protection rating
Programmable analog and switching output
Integrated temperature measurement via IO-Link
Convenient operation via buttons or parameter setting via IO-Link