Найти
Телефон
Меню

Официальный представитель завода АО «Таурас-Феникс»

в Уфе, в Казани, в Оренбурге, в Перми, в Ижевске и в Казахстане

Taures-fenix
+7 347 201 77 10 Бесплатно по России

SV5150 – фото
Фото
Видео

На SV5150

Для заявки пишите в МАХ или Telegram

# #

Характеристики

Output function analog
Materials (wetted parts) ETFE; PA 6T; EPDM
Measuring range 0.265...4.716
Process connection G 3/4 external thread DN10

Все характеристики

Product characteristics
Number of inputs and outputs Number of analog outputs: 1
Measuring range <table><tr><td>1.8...32 l/min</td><td>0.265...4.716 m/s</td></tr></table>
Process connection threaded connection G 3/4 external thread DN10
Application
System gold-plated contacts
Measuring element 1 x Pt 1000; (to DIN EN 60751, class B)
Application for industrial applications
Installation connection to pipe by means of an adapter
Media water; glycol solutions; Coolants
Medium temperature -40...100
Min. bursting pressure <table><tr><td>25 bar</td><td>2.5 MPa</td></tr></table>
Pressure rating <table><tr><td>12 bar</td><td>1.2 MPa</td></tr></table>
Note on pressure rating <table><tr><td>up to 40 °C </td></tr></table>
Electrical data
Operating voltage 8...33 DC
Min. insulation resistance 100; (500 V DC)
Protection class III
Power-on delay time < 2
Measuring principle Vortex
Inputs / outputs
Number of inputs and outputs Number of analog outputs: 1
Outputs
Total number of outputs 1
Output signal analog signal
Number of analog outputs 1
Analog current output 4...20; (water: Q [l/min] = 2,0 x (I - 4 mA); water-glycol: Q [l/min] = 2,0 x (I - 4 mA) - Qo see Figure 2)
Max. load < (Ub - 8 V) / 20 mA; Ub = 24 V: 800
Measuring/setting range
Measuring range <table><tr><td>1.8...32 l/min</td><td>0.265...4.716 m/s</td></tr></table>
Temperature monitoring
Accuracy ± 0,3 ± 0,005 x T
Accuracy / deviations
Flow monitoring
Response time 0.5
Reaction times
Operating conditions
Ambient temperature -15...85
Note on ambient temperature <table><tr><td>medium temperature > 0 °C: -30...85</td></tr></table>
Storage temperature -30...85
Protection IP 65
Cavitation P(absolute) discharge / P(difference) > 5.5 to avoid cavitation
Tests / approvals
EMC <table><tr><td>EN 61326-2-3</td><td></td></tr></table>
Shock resistance <table><tr><td>DIN EN 60068-2-27</td><td>30 g (11 ms)</td></tr></table>
Vibration resistance <table><tr><td>DIN EN 60068-2-6</td><td>with water / 10...61 Hz 1 mm</td></tr><tr><td></td><td>with water / 61...2000 Hz 2 g</td></tr></table>
MTTF 380
Pressure equipment directive sound engineering practice; can be used for group 2 fluids; group 1 fluids on request
Mechanical data
Weight 79
Housing rectangular
Dimensions 90 x 30.22 x 54.1
Material PA 6T
Materials (wetted parts) ETFE; PA 6T; EPDM
Tightening torque 12
Process connection threaded connection G 3/4 external thread DN10
Remarks
Remarks <table><tr><td>MW = Measured value</td></tr><tr><td>MEW = Final value of the measuring range</td></tr></table>
Pack quantity 1 pcs.
Electrical connection - plug
Connection Connector: 1 x M12; coding: A; Contacts: gold-plated

Приемущества

Reliable detection of flow and medium temperature
Precise linearized analog output
Suitable for water and water-based media
For installation in pipes
High precision of flow and temperature monitoring