Найти
Телефон
Меню

Официальный представитель завода АО «Таурас-Феникс»

в Уфе, в Казани, в Оренбурге, в Перми, в Ижевске и в Казахстане

Taures-fenix
+7 347 201 77 10 Бесплатно по России

SV4050 – фото
Фото
Видео

На SV4050

Для заявки пишите в МАХ или Telegram

# #

Характеристики

Output function analog
Materials (wetted parts) ETFE; PA 6T; FKM
Measuring range 0.133...2.21
Process connection G 1/2 external thread DN8

Все характеристики

Number of inputs and outputs Number of analog outputs: 1
Measuring range
0.9...15 l/min0.133...2.21 m/s
Process connection threaded connection G 1/2 external thread DN8
System gold-plated contacts
Measuring element 1 x Pt 1000; (to DIN EN 60751, class B)
Application for industrial applications
Installation connection to pipe by means of an adapter
Media water; glycol solutions; Coolants
Medium temperature -40...100
Min. bursting pressure
25 bar2.5 MPa
Pressure rating
12 bar1.2 MPa
Note on pressure rating
up to 40 °C
Operating voltage 8...33 DC
Min. insulation resistance 100; (500 V DC)
Protection class III
Power-on delay time < 2
Measuring principle Vortex
Number of inputs and outputs Number of analog outputs: 1
Total number of outputs 1
Output signal analog signal
Number of analog outputs 1
Analog current output 4...20; (Q [l/min] = 0,938 x (I - 4 mA))
Max. load < (Ub - 8 V) / 20 mA; Ub = 24 V: 800
Measuring range
0.9...15 l/min0.133...2.21 m/s
Accuracy ± 0,3 ± 0,005 x T
Response time 0.5
Ambient temperature -15...85
Storage temperature -30...85
Protection IP 65
Cavitation P(absolute) discharge / P(difference) > 5.5 to avoid cavitation
EMC
EN 61326-2-3
Shock resistance
DIN EN 60068-2-2730 g (11 ms)
Vibration resistance
DIN EN 60068-2-6with water / 10...61 Hz 1 mm
with water / 61...2000 Hz 2 g
MTTF 380
Pressure equipment directive sound engineering practice; can be used for group 2 fluids; group 1 fluids on request
Weight 65
Housing rectangular
Dimensions 77 x 30.22 x 53
Material PA 6T
Materials (wetted parts) ETFE; PA 6T; FKM
Tightening torque 12
Process connection threaded connection G 1/2 external thread DN8
Remarks
MW = Measured value
MEW = Final value of the measuring range
Pack quantity 1 pcs.
Connection Connector: 1 x M12; coding: A; Contacts: gold-plated

Приемущества

Reliable detection of flow and medium temperature
Precise linearized analog output
Suitable for water and water-based media
For installation in pipes
High precision of flow and temperature monitoring